Abstract
Purpose
Although hydrogel spacer placement (HSP) minimizes rectal dose during prostate cancer radiation therapy, its potential benefit for modulating rectal toxicity could depend on the achieved prostate-rectal separation. We therefore developed a quality metric associated with rectal dose reduction and late rectal toxicity among patients treated with prostate stereotactic body radiation therapy (SBRT).
Methods and Materials
A quality metric consisting of prostate-rectal interspace measurements from axial T2-weighted magnetic resonance imaging simulation images was applied to 42 men enrolled in a multi-institutional phase 2 study using HSP with prostate SBRT (45 Gy in 5 fractions). A score of 0, 1, or 2 was assigned to a prostate-rectal interspace measurement of <0.3 cm, 0.3 to 0.9 cm, or ≥1 cm, respectively. An overall spacer quality score (SQS) was computed from individual scores at rectal midline and ±1 cm laterally, located at the prostate base, midgland, and apex. Associations of SQS with rectal dosimetry and late toxicity were evaluated.
Results
The majority of the analyzed cohort had an SQS of 1 (n = 17; 41%) or 2 (n = 18; 43%). SQS was associated with maximum rectal point dose (rectal Dmax; P = .002), maximum dose to 1 cc of rectum (D1cc; P = .004), and volume of rectum receiving ≥100% of prescription dose (V45; P = .046) and ≥40 Gy (V40; P = .005). SQS was also associated with a higher incidence of (P = .01) and highest-graded late rectal toxicity (P = .01). Among the 20 men who developed late grade ≥1 rectal toxicity, 57%, 71%, and 22% had an SQS of 0, 1, and 2, respectively. Men with an SQS of 0 or 1 compared with 2 had 4.67-fold (95% CI, 0.72-30.11) or 8.40-fold (95% CI, 1.83-38.57) greater odds, respectively, of developing late rectal toxicity.
Conclusions
We developed a reliable and informative metric for assessing HSP, which appears to be associated with rectal dosimetry and late rectal toxicity after prostate SBRT.
Introduction
Given the anatomic proximity of the prostate to normal tissues such as the rectum, acute and late toxicities are possible with prostate cancer external beam or brachytherapy treatments. One such strategy to minimize toxicity is the insertion of a commercially supplied polyethylene glycol hydrogel rectal spacer, a polymerization solution injected transperineally under ultrasound-guidance to temporarily distance the rectum away from the prostate gland. This rectal spacer remains intact throughout a typical course of radiation therapy and is hydrolyzed 3 to 6 months after placement. The first prospective randomized clinical trial to evaluate the effect of the hydrogel rectal spacer in prostate cancer patients treated with conventionally fractionated external beam radiation therapy established its ability to increase the mean prostate-rectal separation by 1 cm; this translated into a lower rectal dose and reduced incidence of late grade 1 to 2 rectal toxicity.
1- Hamstra DA
- Mariados N
- Sylvester J
- et al.
Continued benefit to rectal separation for prostate radiation therapy: Final results of a phase III trial.
,2- Mariados N
- Sylvester J
- Shah D
- et al.
Hydrogel spacer prospective multicenter randomized controlled pivotal trial: Dosimetric and clinical effects of perirectal spacer application in men undergoing prostate image guided intensity modulated radiation therapy.
A systematic review and meta-analysis of 7 studies comparing over 1000 prostate cancer patients treated with dose-escalated external beam radiation therapy and/or brachytherapy with or without hydrogel rectal spacer placement (HSP) reported a 66% V70 rectal dose reduction with a corresponding decline in acute (RR 0.72) and late (RR 0.38) grade 1 and higher rectal toxicity, as well as superior patient-reported >3-month bowel quality of life in the presence of the spacer hydrogel.
3- Miller LE
- Efstathiou JA
- Bhattacharyya SK
- Payne HA
- Woodward E
- Pinkawa M.
Association of the placement of a perirectal hydrogel spacer with the clinical outcomes of men receiving radiotherapy for prostate cancer: A systematic review and meta-analysis.
Attempts have been made to characterize the quality of HSP to accurately assess its true dosimetric- and toxicity-reduction effect given the variability of reported benefit across individual studies. However, these published HSP quality metrics
4- Fischer-Valuck BW
- Chundury A
- Gay H
- Bosch W
- Michalski J.
Hydrogel spacer distribution within the perirectal space in patients undergoing radiotherapy for prostate cancer: Impact of spacer symmetry on rectal dose reduction and the clinical consequences of hydrogel infiltration into the rectal wall.
, 5- Hwang ME
- Black PJ
- Elliston CD
- et al.
A novel model to correlate hydrogel spacer placement, perirectal space creation, and rectum dosimetry in prostate stereotactic body radiotherapy.
, 6- Liu H
- Borden L
- Wiant D
- Sintay B
- Hayes L
- Manning M.
Proposed hydrogel-implant quality score and a matched-pair study for prostate radiation therapy.
have not been universally adopted, as they are somewhat cumbersome to use and were not consistently evaluated for their utility in predicting late rectal toxicity. A secondary analysis of HSP quality in the pivotal phase 3 spacer trial demonstrated that 49% of patients had symmetrical hydrogel rectal spacers in all 3 axial slices analyzed, with a correlation between inferior composite symmetry scores and smaller rectal dose reduction relative to the prespacer plan.
4- Fischer-Valuck BW
- Chundury A
- Gay H
- Bosch W
- Michalski J.
Hydrogel spacer distribution within the perirectal space in patients undergoing radiotherapy for prostate cancer: Impact of spacer symmetry on rectal dose reduction and the clinical consequences of hydrogel infiltration into the rectal wall.
Although informative, this metric does not assess the actual thickness of the prostate-rectal interspace (PRI) created by the spacer gel, which is more relevant to rectal dosimetry.
Because clinical observations have suggested the spacer's ability to reduce rectal dose is dependent on the quality of the implant, a universally adopted practical metric is needed. We therefore created a HSP quality metric that quantitatively characterizes the adequacy of the PRI at the prostate base, midgland, and apex, which also has the benefit of showing association with reductions in rectal dosimetry and late toxicity in men treated with prostate stereotactic body radiation therapy (SBRT).
Discussion
By accounting for the PRI thickness created by the hydrogel rectal spacer at the level of the prostate base, midgland, and apex at midline and 1 cm peripherally, we have created an overall spacer quality score (SQS) that is associated with rectal dosimetry and late rectal toxicity after treatment with SBRT for prostate cancer. Forty-three percent of men had an SQS of 2, highlighting that the highest SQS score is achievable if the insertion of the hydrogel rectal spacer is performed by experienced users in a methodical fashion. An SQS of 2 not only was associated with lower radiation doses to the rectum (as measured by Dmax, D1cc, V45Gy, and V40Gy), but also significantly fewer incidences of late rectal toxicity compared with a score of 0 to 1 (22% vs 67%).
Although previously published HSP metrics have been widely adopted, we present a robust and practical instrument that both predicts for reduced rectal dose and late rectal toxicity. Fischer-Valuck et al
4- Fischer-Valuck BW
- Chundury A
- Gay H
- Bosch W
- Michalski J.
Hydrogel spacer distribution within the perirectal space in patients undergoing radiotherapy for prostate cancer: Impact of spacer symmetry on rectal dose reduction and the clinical consequences of hydrogel infiltration into the rectal wall.
developed a symmetry score in which more asymmetrical implants correlated with less reductions in rectal dose compared with the treatment plan before insertion of the rectal spacer. However, hydrogel rectal spacer symmetry (relative to the prostate midline on axial images) does not always equate to the actual thickness of the PRI in the anterior-posterior direction which, as we demonstrate with our metric, is associated with reduced rectal dose. For example, a patient could have a symmetrical implant, but only minimal (<0.3 cm) space created between the prostate and rectum and therefore derive negligible dosimetric and toxicity benefit. In the Fischer-Valuck et al analysis of HSP quality in the published phase 3 spacer trial,
1- Hamstra DA
- Mariados N
- Sylvester J
- et al.
Continued benefit to rectal separation for prostate radiation therapy: Final results of a phase III trial.
,2- Mariados N
- Sylvester J
- Shah D
- et al.
Hydrogel spacer prospective multicenter randomized controlled pivotal trial: Dosimetric and clinical effects of perirectal spacer application in men undergoing prostate image guided intensity modulated radiation therapy.
all but the most severe asymmetrical implants resulted in decreased rectal dose. This implies that inadequate separation between the prostate and rectum accounted for less dose-sparing to the rectum; less symmetry did result in a smaller rectal dose reduction, however. There was also no attempt by the authors to evaluate association of their symmetry score with rectal toxicity, highlighting the utility of our metric.
Hwang et al
5- Hwang ME
- Black PJ
- Elliston CD
- et al.
A novel model to correlate hydrogel spacer placement, perirectal space creation, and rectum dosimetry in prostate stereotactic body radiotherapy.
used a more complicated model that accounted for implant symmetry in a 3-dimensional orientation and hydrogel rectal spacer volume. However, they did not identify an association between their metric and rectal toxicity in men treated with prostate SBRT.
5- Hwang ME
- Black PJ
- Elliston CD
- et al.
A novel model to correlate hydrogel spacer placement, perirectal space creation, and rectum dosimetry in prostate stereotactic body radiotherapy.
The authors attributed this potentially to a small sample size of 22 patients, the moderate SBRT dose of 36.25 Gy delivered in 5 fractions, and inadequate duration of follow-up. However, we observed a similar rate of late grade 1 to 2 rectal toxicity (48% in our vs 30% in the Hwang et al studies) in a small sample size of 42 men that was still associated with late rectal toxicity. A mathematical equation consisting of 4 component equations (hydrogel volume, symmetry in the left-right and cranial-caudal orientations, and prostate-rectal space at midgland) was derived by Liu, et al,
6- Liu H
- Borden L
- Wiant D
- Sintay B
- Hayes L
- Manning M.
Proposed hydrogel-implant quality score and a matched-pair study for prostate radiation therapy.
to quantify HSP quality; no meaningful associations, however, were correlated with this score as well.
Increasing SBRT doses delivered to the prostate have been correlated with lower 2-year posttreatment positive biopsy rates
9- Zelefsky MJ
- Kollmeier M
- McBride S
- et al.
Five-year outcomes of a phase 1 dose-escalation study using stereotactic body radiosurgery for patients with low-risk and intermediate-risk prostate cancer.
and superior biochemical control rates.
10- Jackson WC
- Silva J
- Hartman HE
- et al.
Stereotactic body radiation therapy for localized prostate cancer: A systematic review and meta-analysis of over 6,000 patients treated on prospective studies.
Thus, such measures used to attenuate dose to the rectum by anatomically separating the prostate and rectum via an implantable hydrogel rectal spacer are even more important. In a retrospective study from MSK of 551 men treated with SBRT for prostate cancer, those who had hydrogel rectal spacers implanted before treatment developed fewer late grade 2 and higher rectal toxicity (1%) compared with those without an implant (6%).
11- Zelefsky MJ
- Pinitpatcharalert A
- Kollmeier M
- et al.
Early tolerance and tumor control outcomes with high-dose ultrahypofractionated radiation therapy for prostate cancer.
Because studies using the hydrogel rectal spacer have not reported on the quality of the implant, it will be difficult to completely understand its benefits if a uniform QA metric is not widely used. Poor implants, which would be defined by an SQS of 0, will unnecessarily have placed a patient at risk of complications from the procedure without providing a reduced risk of developing toxicity from the actual treatment. Our SQS metric has the benefit of quantifying the total space between the prostate and rectum, which includes both the natural anatomic space and the additional space created by the hydrogel implant. It has been previously reported that fewer postbrachytherapy rectal complications occur with larger natural space present between the prostate and rectum.
12- Kang MH
- Yu YD
- Shin HS
- Oh JJ
- Park DS.
Difference in the rate of rectal complications following prostate brachytherapy based on the prostate-rectum distance and the prostate longitudinal length among early prostate cancer patients.
Therefore, in men who have favorable anatomy in certain areas along the PRI, it could potentially be decided before implantation where the hydrogel rectal spacer should be focally inserted to enhance this separation.
Furthermore, we anticipate SQS's utility in not only reporting late rectal toxicity with external beam and brachytherapy prostate treatments in the presence of HSP, but also as a self-improvement tool for radiation oncologists to identify areas for refinement of the implant technique; visualization of the scores within the 3 × 3 table facilitates assessment of spacer symmetry both in the left-right and superior-inferior orientations. We anticipate that physicians who are more experienced with placement of the hydrogel rectal spacer will, on average, have implants with a higher SQS. For example, Pinkawa et al
13- Pinkawa M
- Klotz J
- Djukic V
- et al.
Learning curve in the application of a hydrogel spacer to protect the rectal wall during radiotherapy of localized prostate cancer.
assessed a group of 64 consecutive patients with rectal hydrogel spacers inserted at their institution over time and demonstrated that a learning curve existed: the latter half of patients had improved spacer symmetry, larger prostate-rectal separations, moderated rectal dose, and fewer acute rectal toxicity after radiation therapy.
A limitation of this study includes the small sample size, which may have weakened our ability to determine whether the SQS should be reported as a 3-tier or binary score (in which 0 and 1 are grouped together). Furthermore, analysis of late toxicity did not account for follow-up time because this data was unavailable; therefore, we acknowledge that the results may be biased. Furthermore, we do not have data to assess whether SQS predicts which patients will have resolution of their late rectal toxicity. Another weakness of this study is that most patients who developed a late rectal toxicity had a grade of 1 (14 of 20 patients with toxicity), which we recognize as generally considered clinically insignificant. However, prostate SBRT, especially when a hydrogel rectal spacer is present, is a well-tolerated treatment with <5% late grade 2 and higher rectal toxicity rates.
11- Zelefsky MJ
- Pinitpatcharalert A
- Kollmeier M
- et al.
Early tolerance and tumor control outcomes with high-dose ultrahypofractionated radiation therapy for prostate cancer.
Similarly, there was no grade 2 or higher late rectal toxicity in men with HSP treated with conventionally fractionated radiation to the prostate in the pivotal phase 3 spacer study.
1- Hamstra DA
- Mariados N
- Sylvester J
- et al.
Continued benefit to rectal separation for prostate radiation therapy: Final results of a phase III trial.
,2- Mariados N
- Sylvester J
- Shah D
- et al.
Hydrogel spacer prospective multicenter randomized controlled pivotal trial: Dosimetric and clinical effects of perirectal spacer application in men undergoing prostate image guided intensity modulated radiation therapy.
Therefore, it is reasonable that our SQS metric can predict any graded late rectal toxicity. All of these shortfalls highlight the necessity to externally validate the SQS using a larger sample size, with more complete toxicity data, and with different radiation therapy techniques (ie, SBRT, moderate hypofractionation, and brachytherapy) to ensure its versatility and utility.
Although this comprehensive instrument that accounts for hydrogel spacer symmetry and thickness throughout the entire irradiated duration of the prostate-rectal interface was associated with both rectal dosimetry and late rectal toxicity in this small sample size, a more simplistic metric might achieve the same goals. Using the longer follow-up data from a published prospective randomized trial that compared rectal toxicity and dosimetry in patients treated with definitive radiation in the presence or absence of a hydrogel rectal spacer,
1- Hamstra DA
- Mariados N
- Sylvester J
- et al.
Continued benefit to rectal separation for prostate radiation therapy: Final results of a phase III trial.
we plan to rigorously test the validity of the SQS and compare it to reduced metrics that describe the quality of separation at only one level of the prostate gland.
Article info
Publication history
Published online: January 23, 2023
Accepted:
September 1,
2022
Received:
January 19,
2022
Footnotes
Sources of support: This work had no specific funding.
Disclosures: Drs Folkert and Zelefsky are consultants for Boston Scientific Inc. No other disclosures were reported.
Research data from this work are stored in an institutional repository and will be shared upon request to the corresponding author.
Copyright
© 2023 The Authors. Published by Elsevier Inc. on behalf of American Society for Radiation Oncology.